How to use the DAC curve calculator
- Pick the position unit. Use sound path or depth for every point and the indication, not a mix of the two.
- Enter 2 to 6 reference points: where each reference echo peaked and its height in percent full-screen height (% FSH), all at the reference gain. Order does not matter.
- Enter the indication's position and its peak height as read on the screen.
- If you added or removed gain to read the indication, enter the change in dB: positive for more gain than the reference, negative for less.
The formulas
- The curve between two points. DAC(s) = A1 + (s − s1) ÷ (s2 − s1) × (A2 − A1), for a position s between neighboring points s1 and s2. The calculator joins the marks with straight lines, as the DAC curve guide describes the curve being drawn, and never extends it past the first or last point.
- Back to the reference gain: A' = A × 10^(−ΔG ÷ 20). Every 6 dB of extra gain roughly doubles an echo, so an echo read 6 dB hot is about half as tall at the reference gain.
- dB re DAC = 20 × log10(A' ÷ DAC(s)) and %DAC = 100 × A' ÷ DAC(s). The 20 × log10 form for two screen heights is the decibel definition in Evident's flaw detection tutorial glossary.
- −6 dB and −14 dB lines: DAC × 10^(−6 ÷ 20) and DAC × 10^(−14 ÷ 20), about 50.1% and 20.0% of the curve. Evident's tutorial shows these two extra curves below a reference curve as an example. Which lines your procedure uses, if any, is up to the procedure.
How the reference points are made
Evident's tutorial says DAC is usually set up with a series of reference reflectors of one size at different depths, from a bought distance/amplitude block set or a block users make themselves. With the probe and instrument settings fixed, peak each reflector in turn and note its position and height. Those pairs are the points you enter here. The curve belongs to that probe, those settings and that material, so a change to any of them means a new curve.
Worked example 1: an echo on a reference point
Invented numbers. Points at 10, 20, 30 and 40 mm read 80%, 64%, 50% and 40% FSH. An indication at 20 mm peaks at 64% FSH with no gain change. The curve there is 64%, so dB re DAC = 20 × log10(64 ÷ 64) = 0 dB and the echo is 100% DAC.
Worked example 2: half the curve height
Same points. At 30 mm the curve is 50% FSH and an indication peaks at 25% with no gain change. dB re DAC = 20 × log10(25 ÷ 50) = −6.02 dB, or 50% DAC. Halving an amplitude is always about −6 dB.
Worked example 3: between two points, read with extra gain
Same points. An indication at 25 mm peaks at 60% FSH, read 6 dB above the reference gain. The curve at 25 mm is 64 + (25 − 20) ÷ (30 − 20) × (50 − 64) = 57% FSH. At the reference gain the echo is 60 × 10^(−6 ÷ 20) = 30.07% FSH. dB re DAC = 20 × log10(30.07 ÷ 57) = −5.55 dB, or 52.8% DAC. Forget the gain change and you would rate the echo 1.05 times the curve, above it instead of below.
Move that indication to 45 mm and the calculator gives no rating, because the last point is at 40 mm and the curve is not extended. Add a reference point at or beyond 45 mm instead.
These are original arithmetic examples, not exam questions. The same decibel steps are in the 6 dB and 20 dB rules, and the decibel calculator converts any ratio.
Common mistakes
- Wrong sign on the gain change. Extra gain makes the echo taller, so it must come off: A' = A × 10^(−ΔG ÷ 20).
- Reading a clipped echo. An echo above 100% FSH has no true height. Drop the gain, read it on screen and enter the gain change.
- Extending the curve. Past the last point you have no reference echo. Add one rather than guess.
- Mixing sound path and depth. Points and the indication must use the same measure.
- Using a stale curve. A new probe, couplant, gain setting or material means rebuilding or verifying the curve. Attenuation and the near field explain why it changes.
DAC curve questions
What is a DAC curve in ultrasonic testing?
DAC stands for distance amplitude correction. The curve joins the echo heights from reflectors of one size at several distances from the probe. The same reflector gives a lower echo farther away, mostly from attenuation and beam spread, so the curve slopes down. Evident's flaw detection tutorial says an echo from a reflector the same size as the reference ones matches the curve height at any depth, and a larger or smaller one sits above or below it.
How do you work out dB above or below DAC?
Bring the echo to the reference gain first: A' = A × 10^(−ΔG ÷ 20), where ΔG is how many dB above the reference gain it was read at. Then dB re DAC = 20 × log10(A' ÷ DAC), using the curve height at the echo's own position. An echo at half the curve height is about −6 dB, and %DAC = 100 × A' ÷ DAC.
Does this calculator say whether to record or reject an indication?
No. Your written procedure sets the reference level, how the curve is built and any recording or acceptance level, and a qualified person makes the call. This page does the arithmetic only and holds no acceptance level from any code or standard.
Related: the UT Level II formulas guide for the rest of the UT arithmetic, and the UT beam calculator for the near field and beam spread behind the curve's shape.